Case Study 3: Gaussian Laser Beam Expander

Goal

This case study demonstrates the laser workflow in the UI:

  • enter manufacturer-style laser data: beam diameter and full divergence;

  • let KrakenOS back-calculate the waist radius and waist location;

  • inspect the Gaussian q propagation report;

  • convert a free-space laser line into a 3x Keplerian beam expander;

  • verify the before/after beam size and divergence;

  • run the BField analysis to show detector-plane field intensity and phase.

The screenshots in this tutorial are generated from the live Tk UI with:

python -m KrakenOS.UI.capture_gaussian_beam_expander_case_study_screenshots

Load The Laser Line

  1. Start the UI with python -m KrakenOS.UI.layout_editor.

  2. In the top menu choose Layouts -> Sources / Illumination -> Gaussian Laser Beam Expander Case Study.

  3. Confirm the Source panel uses laser-datasheet input:

Source model       = Gaussian beam
GB input mode      = Diameter + divergence
GB diameter [mm]   = 1.0
GB full div [mrad] = 1.0
GB waist side      = Waist before source
GB M2              = 1.0
Wavelength [um]    = 0.6328

This means the laser output plane has a 1/e² beam diameter of 1.0 mm and a full far-field divergence of 1.0 mrad. The UI hides or disables the ordinary field/pupil inputs because the physical Gaussian source now defines the illumination.

Gaussian source panel configured with diameter and divergence

The Source panel is in Gaussian beam mode. Diameter and divergence are the active input fields; object-field controls are not part of this source workflow.

Click Update. The 2D layout shows the traced representative source rays and the amber 1/e² Gaussian envelope.

Free-space Gaussian laser layout with 1/e squared envelope

Free-space baseline: a 1 mm diameter, 1 mrad full-divergence laser propagates to the readout plane.

Back-Calculated Waist

Open Actions -> Gaussian Beam Report.

For this laser input, KrakenOS computes approximately:

waist radius w0       ~= 0.403 mm
waist location offset ~= 592 mm before the source plane

At the free-space readout plane, the beam is larger but the divergence is still the original datasheet value:

final 1/e^2 diameter ~= 1.36 mm
final half divergence ~= 0.50 mrad
Gaussian Beam Report for free-space laser baseline

The Gaussian Beam Report lists the ABCD step, q value, beam radius, wavefront radius, waist offset, Rayleigh range, divergence, and Gouy phase.

Insert A 3x Keplerian Expander

The target is a 3x Keplerian expander. Use two positive lenses separated by f1 + f2:

magnification = f2 / f1 = 150 / 50 = 3
lens separation = f1 + f2 = 200 mm

Edit the table so it has these rows:

Row

Surface

Name

Rc [mm]

Thickness [mm]

Diameter [mm]

Material

0

Object

Laser output

0

80

16

AIR

1

Thin Lens

Input lens f=50

50

200

20

AIR

2

Thin Lens

Collimating lens f=150

150

320

45

AIR

3

Image

Readout plane

0

0

50

AIR

Practical click path:

  1. Change row 0 Thickness from 500 to 80.

  2. Click Add surface twice. New rows are inserted before Image.

  3. For each inserted row, right-click the Surface cell and choose Thin Lens.

  4. Fill the names, Rc focal lengths, thicknesses, and diameters from the table above. The UI will also show an automatic Src1 illumination-source row between Object and the optical surfaces; that row is expected and is not one of the KrakenOS surface rows you are editing.

  5. Click Update.

Gaussian beam expander table after inserting two thin lens rows

The table now contains the Keplerian expander: f1=50 mm followed by f2=150 mm.

Thin Lens rows are ideal ABCD lens elements. In the 2-D layout they are drawn with a purple/orange lens glyph for readability; the glyph width is not a physical center thickness.

Verify The Expanded Beam

The output beam should be about three times wider and about three times less divergent than the input beam. Click Update and inspect the Gaussian envelope.

Gaussian beam envelope after Keplerian expander

The amber envelope expands through the positive/positive lens pair and then propagates with lower divergence.

Open Actions -> Gaussian Beam Report again. The final readout-plane values should be close to:

final 1/e^2 diameter ~= 3.09 mm
final half divergence ~= 0.167 mrad

This is the expected 3x beam expansion and 3x divergence reduction.

Gaussian Beam Report after Keplerian beam expander

The report verifies the q-parameter propagation through both thin lenses. The final divergence is about one third of the original value.

Run BField Analysis

The Gaussian q report is the paraxial design check. The BField button is the detector-field check:

  1. Set Detector bins = 64.

  2. Set Coherent sum = Mutual coherent.

  3. Set BField z [mm] = 0.

  4. Click BField and then Update.

BField detector intensity and phase for the expanded Gaussian beam

BField promotes the detector samples into the branch-field grid and plots normalized field intensity, phase contours, centroid, and TEM00 overlap diagnostics.

What This Proves

This case study exercises the laser-specific UI path:

  • Gaussian source model with physical source position and direction;

  • laser-datasheet input mode;

  • automatic waist back-calculation;

  • 1/e² Gaussian envelope overlay in the 2D layout;

  • Gaussian Beam Report / ABCD q propagation;

  • thin-lens beam-expander design;

  • detector-side branch-field analysis.

Common Mistakes

I entered half-angle divergence.

The UI field is full divergence in milliradians. A 1 mrad full-angle laser uses GB full div [mrad] = 1.0.

The ordinary Field panel disappeared.

That is intentional. For a physical Gaussian source, the source position, direction, diameter, divergence, and power define the illumination.

The Gaussian envelope is not shown.

The amber q-envelope overlay is for centered +Z paraxial layouts. For tilted, folded, or beam-splitter layouts, use traced rays plus Gaussian Beam Report / Branch Gaussian Q Report.

The expander output is bigger, so did it make divergence worse?

No. A beam expander increases beam diameter and reduces angular divergence. In this example the diameter is about 3x larger and the divergence is about 3x smaller.